Disassemblable Fieldwork CT Scanner Using a 3D-printed Calibration Phantom
원문 초록
The use of computed tomography (CT) imaging has become of increasing interest to academic areas outside of the field of medical imaging and industrial inspection, e.g., to biology and cultural heritage research. The pecularities of these fields, however, sometimes require that objects need to be imaged on-site, e.g., in field-work conditions or in museum collections. Under these circumstances, it is often not possible to use a commercial device and a custom solution is the only viable option. In order to achieve high image quality under adverse conditions, reliable calibration and trajectory reproduction are usually key requirements for any custom CT scanning system. Here, we introduce the construction of a low-cost disassemblable CT scanner that allows calibration even when trajectory reproduction is not possible due to the limitations imposed by the project conditions. Using 3D-printed in-image calibration phantoms, we compute a projection matrix directly from each captured X-ray projection. We describe our method in detail and show successful tomographic reconstructions of several specimen as proof of concept.
함께 살펴볼 논문
공통 태그를 근거로 연결했습니다. 실제 인용 관계나 논문의 우수성을 뜻하지 않습니다.
Enhancing Synthetic CT from CBCT via Multimodal Fusion and End-To-End Registration
SAM2-SGP: Enhancing SAM2 for Medical Image Segmentation via Support-Set Guided Prompting
TissUnet: Improved Extracranial Tissue and Cranium Segmentation for Children through Adulthood
pyMEAL: A Multi-Encoder Augmentation-Aware Learning for Robust and Generalizable Medical Image Translation
Domain and Task-Focused Example Selection for Data-Efficient Contrastive Medical Image Segmentation
Accelerating Volumetric Medical Image Annotation via Short-Long Memory SAM 2
토론
이 논문과 연결된 포럼 글이 아직 없습니다.