High quality ultrasonic multi-line transmission through deep learning
원문 초록
Frame rate is a crucial consideration in cardiac ultrasound imaging and 3D sonography. Several methods have been proposed in the medical ultrasound literature aiming at accelerating the image acquisition. In this paper, we consider one such method called \textit{multi-line transmission} (MLT), in which several evenly separated focused beams are transmitted simultaneously. While MLT reduces the acquisition time, it comes at the expense of a heavy loss of contrast due to the interactions between the beams (cross-talk artifact). In this paper, we introduce a data-driven method to reduce the artifacts arising in MLT. To this end, we propose to train an end-to-end convolutional neural network consisting of correction layers followed by a constant apodization layer. The network is trained on pairs of raw data obtained through MLT and the corresponding \textit{single-line transmission} (SLT) data. Experimental evaluation demonstrates significant improvement both in the visual image quality and in objective measures such as contrast ratio and contrast-to-noise ratio, while preserving resolution unlike traditional apodization-based methods. We show that the proposed method is able to generalize well across different patients and anatomies on real and phantom data.
함께 살펴볼 논문
공통 태그를 근거로 연결했습니다. 실제 인용 관계나 논문의 우수성을 뜻하지 않습니다.
Scaling Laws of Dynamic High-Capacity Ride-Sharing
CUTS+: High-dimensional Causal Discovery from Irregular Time-series
CosmoPower-JAX: high-dimensional Bayesian inference with differentiable cosmological emulators
HumanRF: High-Fidelity Neural Radiance Fields for Humans in Motion
MIReAD: Simple Method for Learning High-quality Representations from Scientific Documents
High-Fidelity 3D Face Generation from Natural Language Descriptions
토론
이 논문과 연결된 포럼 글이 아직 없습니다.