Measuring uncertainty during respiratory rate estimation using pressure-sensitive mats
원문 초록
We develop and evaluate a respiratory rate estimation algorithm that utilizes data from pressure-sensitive mat (PSM) technology for continuous patient monitoring in neonatal intensive care units (NICU). An analysis of the random effect of drift and systematic effect of creep in the PSM data is presented, showing that these are essentially dependent on the applied load and contact surface. Uncertainty measurements are pivotal when estimating physiologic parameters. The standard uncertainty in the PSM data is here represented by the percent drift. Next, we evaluate the applicability of PSM technology to estimate RR in neonatal patient simulator trials under five mixed effects including internally and externally induced motion, mattress type, grunting, laying position, and different breathing rates. We analyze the limits of agreement on the mixed effects model to derive the uncertainty in the estimated RR obtained through two estimation techniques. In comparison with the gold standard RR values, we achieved a mean bias of 0.56 breaths per minute (bpm) with an error bounded by a 95% confidence interval of [-2.26, 3.37] bpm. These results meet the clinical accuracy requirements of RR within +/-5 bpm.
함께 살펴볼 논문
공통 태그를 근거로 연결했습니다. 실제 인용 관계나 논문의 우수성을 뜻하지 않습니다.
Automatic Infant Respiration Estimation from Video: A Deep Flow-based Algorithm and a Novel Public Benchmark
VitalLens: Take A Vital Selfie
RRWaveNet: A Compact End-to-End Multi-Scale Residual CNN for Robust PPG Respiratory Rate Estimation
Radar-based Respiratory Rate Monitoring in Standing Position
Dual Attention Network for Heart Rate and Respiratory Rate Estimation
Respiratory Rate Estimation Based on WiFi Frame Capture
토론
이 논문과 연결된 포럼 글이 아직 없습니다.